Precautions for Vacuum Pressure Gauge

1. The function of vacuum pressure gauge

The vacuum pressure gauge is suitable for measuring the pressure of liquid, steam, gas and other media that are non-explosive, non-crystallized, non-solidified, and non-corrosive to steel and copper alloys. It is generally used in the vacuum measurement of industrial processes such as petroleum, chemical, metallurgy and so on.


2. Precautions for using vacuum pressure gauge

1) The meter must be vertical: a 17mm wrench should be used for tightening during installation, and the case should not be forcibly twisted; collisions should be avoided during transportation;
2) The ambient temperature of the vacuum pressure gauge should be -25-55℃;
3) The vibration frequency of the working environment is less than 25HZ, and the amplitude is less than 1mm;
4) If the indicator value does not return to zero or the indicator value is out of tolerance due to the high ambient temperature during use, you can cut the sealing rubber plug on the upper part of the case to make the inner cavity of the instrument communicate with the air;
5) The use range of vacuum pressure gauge should be between 1/3 and 2/3 of the upper limit;
6) Isolation devices should be added when measuring corrosive media, media that may crystallize, and media with high viscosity;
7) The meter should be inspected frequently (at least once every three months), and if any malfunction is found, it should be repaired in time;
8) If an instrument for measuring corrosive media is required, the required conditions should be indicated when ordering.


The above is the precautions for the use of the vacuum pressure gauge. The vacuum pressure gauge is an indispensable test instrument when repairing refrigeration equipment.

Related Article for Reference

Installation Procedure of Pressure Gauges
The measuring system is essential for pressure gauging. Correct selection and verification (calibration) of the pressure gauge are very important for accurate measurement. Besides, correct installatio...
Thu 11 2020
Problems Related to the Field Test of SF6 Gas Density Monitor
In the SF6 gas density monitor circuit breaker, the decisive factor for its electrical strength and arc extinguishing performance is the gas density, and the main component reflecting the change of th...
Tue 05 2021
Verification of SF6 Gas Relay at Extreme Temperatures
Currently, the SF6 relay is mainly calibrated at room temperature, and it is unable to accurately evaluate the temperature of the relay under extreme conditions. When the environmental temperature bec...
Wed 08 2023
Lanso Instruments INC.
233 W 1st St #210, North Vancouver, Canada
info@lansoinstruments.com
233 W 1st St #210, North Vancouver, Canada
info@lansoinstruments.com